48
SERVICE
6. Push the needle through the upper adapter toward the septum of the glycerin vial.
7. Press the sensor through the glycerin vial septum (
).
Be sure that the sensor is inserted completely and no air bubbles are produced near
the sensor.
Glycerin vial
Upper adapter
Lower adapter
Sensor
Figure 48 Glycerin file with sensor inserted
8. Open the LABROS SoilView software.
9. Verify that the devices are connected.
10. Configure the measurement by entering a sample, file name, and selecting the
measurement profile for verification.
The recommended and preset parameters are shown in