49
VARIOS
Figure 49 Example VARIOS Profile Verification Glycerin
11. Start the measurement as described in
.
NOTE: If the vial is moved or shaken while the instrument is measuring the glycerin verification standard, the
readings will be adversely impacted.
The measurement stops automatically after recording 10 data points.
The software automatically verifies if the recorded values are in the defined range for a
successful verification and delivers a corresponding message. Measurements must be
within ±5 % of the glycerin standard.
NOTE: After repeated use, the septum may begin to leak. To replace, simply remove the cap, press the old septum out,
and replace with a new one.
4.2 CLEANING
The stainless-steel needles can be cleaned with steel wool or other mild abrasives. The
needles can also be straightened with a pair of pliers by applying gentle pressure, if needed.
CAUTION
Be very gentle when bending the needles to avoid damaging the heating element and temperature sensor in the needle.
The VARIOS can be cleaned with a paper towel dampened with water.
NOTE: The VARIOS case and connector plugs are not waterproof. Therefore, DO NOT clean the VARIOS under
running water. DO NOT clean the device with other substances such as alcohol to avoid damaging the VARIOS
case material.