4
OPERATION
Table 1 Installation (continued)
Installation
Install LABROS SoilView Software and METER USB Driver
Download the
LABROS SoilView Software Download
.
Run the
LABROS SoilView.exe
file and follow the installation instructions.
NOTE: LABROS SoilView software can only be used with Microsoft Windows 10 or newer.
If there is a problem executing the LABROS SoilView.exe, please download and
install Microsoft.NET Framework 4.8.0 manually and have the current version
of the Microsoft Windows installer.
Click on the LABROS SoilView icon to open the program.
Connect VARIOS to HYPROP Balance (VARIOS ready)
Insert the VARIOS connector cable into the back of the HYPROP Balance
(
).
Connect VARIOS to a HYPROP Balance That is not VARIOS Ready
Connect the VARIOS connector cable into the VARIOS power adapter, and plug
the power adapter into the HYPROP Balance (
).
When the VARIOS is powered, a blinking white light will appear on the front of
the unit. See
for more information regarding the lights and meanings.
2.2 KEY FACTORS TO GET RELIABLE MEASUREMENTS
Getting reliable results from VARIOS measurements is dependent upon the following factors.
• Air gaps in the sample may form if the needle moves around in the sample.
If an air gap forms, the accuracy of the measurement will be affected. The sensor needle
needs to fit tightly in the sample to reduce the chance of the needle moving around.
• Small stones or pebbles in the sample may prevent the sensor needle from being inserted
into the sample.
Try installing the sensor through the second drilling hole.
2.3 CONNECT VARIOS CONFIGURATIONS
The VARIOS can be configured two ways depending on the type of measurement data
required and the accessories available. This section describes the two configurations and
how to connect the accessories.
2.3.1 VARIOS THERMAL DRYOUT CONFIGURATION
The VARIOS Thermal dryout configuration uses both the VARIOS and a HYPROP Balance to
calculate the thermal conductivity as a function of water content of soil samples or various
suitable materials. Both the VARIOS Complete Set and the Extension Set can be used for the
Thermal dryout configuration.