24
OPERATION
Table 2 HYPROP Balance Settings
Settings
Display
Adjustment
Bar graph
1. B.G. 1
1 = on
Limit function
2. SEL 0
0 = off
Auto-zero
3. A.0 1
1 = on
Response speed
5. rE. 3
3 = medium speed
Stability parameters
6. S.D. 2
2 = medium sensitivity
Interface
7. I.F. 1
1 = 6-digit numeric format
• Select a setting by pressing the
FUNCTION button.
• Change the setting value by using the
ZERO/TARE button.
• Terminate the function selection by pressing the
SET button.
lists possible error messages and troubleshooting solutions.
2.7.4 CONFIGURE A MEASUREMENT
This section contains information that describes how to configure a soil sample
measurement using the LABROS SoilView software.
Remove the measurement pan and place the VARIOS unit on the HYPROP Balance. Place the
soil sample ring on the VARIOS, centered on the mark on top of the VARIOS.
Figure 25 Configure a measurment
To start a measurement do the following:
1. Enter a sample name.
2. Select the type of sample ring (250 mL or customized).
NOTE: A custom soil container can be used with the VARIOS. The container volume and weight will need to be
recorded and entered into the LABROS SoilView-Analysis software. The maximum weight that the balance can
read is 2,200 g.
3. Choose a folder where the data will be saved.
4. Select the Measurement Profile defined in the settings.