data handling:
These settings determine the action of the device in
case of the measurement is to long for screening (the
data area is made for 2048 readings).
Pressing the
ENTER
key twice gives you two options:
1.
"if
data area overflow: no change but send to PC“
New readings will not be shown anymore. However, if a
PC is connected the readings will still be sent to the PC.
2.
"if
data
area
overflow
cut
oldest
values“
For each new reading the oldest reading will be
overwritten and shifted to the first position in the buffer
(ring buffering).
If the option "auto save“ was chosen (see chapter 12.2
or 12.3, Setting the Measuring Parameters and Display
Options) readings of the overflown data area will be
saved into the device memory.
value log.:
kind of getting the readings
“
direct
” One bubble will be used for getting one reading.
“
adapted
” In case of lower bubble lifetimes several
bubbles will be used for getting one reading.
language helptext
: Changing the language of the on-line help.