By using the cursor keys 

  

 the x-axis (bubble lifetime) can be zoomed 

in and out. 

 

The following measuring parameters and display options can be set (saving 

in case of starting a new measurement): 

control:

 

Either  the  bubble  "lifetime"  or  the  "frequency"  can  be 

controlled.  (For  setting  the  control  to  "frequency"  assume 

"bubble  frequency"  instead  of  "bubble  lifetime"  in  the 

following explanation.) 

bubble lifetime

 

 

/ bubble frequency

 

start:

 

Start value of bubble lifetime / frequency. 

stop: 

End value of bubble lifetime / frequency. 

factor: 

Determines the step by which the start value will be changed 

until  the  end  value    was  reached. For each step the current 

lifetime / frequency will be multiplied by this factor. 

auto clr data: 

Clear  the 

shown  data  (diagram  or  table) 

by  starting  a  new 

measurement („on“ / “off“) 

autosave:

 

Save  the 

shown  data  (diagram  or  table) 

into  the  device 

memory after the measurement is finished ("on" / "off"). It is 

useful to set “

auto clr data

” on. 

graphic settings... 

 

Please, refer to 12.2 "On-line Measur

ement“. 

table settings...

 

 

 

 

Please, refer to 12.2 "On-

line Measurement“. 

Summary of Contents for SITA science line t60

Page 1: ...S I T A s c i e n c e l i n e t 6 0 M a n u a l...

Page 2: ...k GmbH 1 Contents 1 CONTENTS 2 2 SITA ONLINE T60 4 3 TECHNICAL DATA 5 4 CONNECTORS 6 5 SETTING UP THE DEVICE 7 6 INSERTING OR CHANGING THE BATTERY 7 7 CAPILLARY AND TEMPERATURE SENSOR 8 8 DRYING AGENT...

Page 3: ...NG CONTRAST AND BRIGHTNESS 13 12 MODES 14 12 1 CALIBRATION 14 12 2 ON LINE MEASUREMENT 15 12 3 AUTO MEASUREMENT 18 12 4 CLEANING MODE 20 13 MEMORY AND DEVICE PARAMETERS 20 13 1 MEMORY 20 13 2 SETTING...

Page 4: ...done by pumping air through a capillary with the radius r into the liquid to be analysed According to Young Laplace the difference in pressure p which is needed to form a bubble is proportional to the...

Page 5: ...illary Connector Luer Lock Connector Surface Tension Measuring Range 10 100 mN m dyn cm Error max 1 f s Resolution 0 1mN m Repeatability Error 0 5mN m Temperature of Liquid Measuring Range 10 130 C Er...

Page 6: ...6 SITA Messtechnik GmbH 4 Connectors 1 Drying agent exchangeable 2 Socket for pneumatic cable to the sensor head 3 Socket for PC connection 4 Socket for power supply...

Page 7: ...o mains or insert 4 mignon AA batteries into the battery compartment watch for the right polarisation Additionally you may want to connect the device to the serial port of your PC This can be done by...

Page 8: ...mply plug it onto the Luer connector and turn to the right Please keep the inner seal clean in order to guaranty a sealed system Note that the PEEK capillaries are not resistant against concentrated s...

Page 9: ...ary The drying agent in the cartridge absorbs most of the moisture lowering the dew point The dew point marks the beginning of condensation The desiccant cartridges come in lockable air tight bags Whe...

Page 10: ...10 SITA Messtechnik GmbH 9 Keyboard...

Page 11: ...echnik GmbH For turning the device on and off hold down for 1 second 2x Start on line measurement 2x Start auto measurement Start cleaning mode Call menu Escape Confirm Cursor keys Display illuminatio...

Page 12: ...on on the measurement and options and the current function of the software keys are shown in different areas of the display The Status is displayed in the first line and second to the last line of the...

Page 13: ...enu button and chose device set up by pressing the arrow keys and confirming with the ENTER button From the menu which is displayed now chose display setting and press ENTER The options contrast and b...

Page 14: ...the table in Surface Tension of Water For calibrating the tensiometer the sensor head s capillary and temperature sensor must be immersed in water Lower the sensor head until the water surface is wit...

Page 15: ...n these two modes by pressing the middle software key table graph When in online mode the surface tension will be measured at the set frequency Make sure that the tensiometer has already been calibrat...

Page 16: ...set takes you to the set up menu In order to continue the measurement you need to press the middle software key go on By using the cursor keys you can scroll the x axis number of reading The followin...

Page 17: ...ermines at which position new readings will be inserted into the table at the top or appended at the end show value of Defines which measured value is to be shown the bubble lifetime t life the bubble...

Page 18: ...ou may check or change the parameters of the measurement to be started You can start the auto measurement from the menu too Press the menu key and select auto measure using the cursor keys Select the...

Page 19: ...quency start Start value of bubble lifetime frequency stop End value of bubble lifetime frequency factor Determines the step by which the start value will be changed until the end value was reached Fo...

Page 20: ...etone This can heavily damage the glued capillary assembly and so causing air leakage Press the clean key to start the cleaning procedure You can also start the cleaning by choosing CLEANING from the...

Page 21: ...essing the ENTER key an input mask will come up showing the current values Navigate to the desired position using the arrow keys and adjust the value using the other pair of arrow keys Beeper The alar...

Page 22: ...data area overflow cut oldest values For each new reading the oldest reading will be overwritten and shifted to the first position in the buffer ring buffering If the option auto save was chosen see...

Page 23: ...ontrol lifetime Values p lifetime 1 pts Start 1 sec Stop 10 sec Factor 1 5 auto clr data off auto save off The values of the graphic settings and table settings will not be reset An alternative to pre...

Page 24: ...ion of device too high battery low Battery low temp device low Device temperature below 0 C temp device high Device temperature above 50 C temp sensor low Temperature sensor head below 0 C temp sensor...

Page 25: ...se read the software licence agreement prior to opening the software package s or using the software The licence agreement is contained in this manual or is attached to the software package By opening...

Page 26: ...ed surface tension maximum of 128 measurements in one window plotting the measurements versus time number of measurements temperature bubble frequency or bubble lifetime b saving the plots Note The so...

Page 27: ...74 69 32 70 86 57 66 67 82 62 10 8 74 54 33 70 70 58 66 49 83 61 91 9 74 39 34 70 54 59 66 31 84 61 71 10 74 25 35 70 38 60 66 14 85 61 52 11 74 10 36 70 22 61 65 96 86 61 33 12 73 95 37 70 05 62 65...

Page 28: ...reaty provisions You must treat the Software as you would any other copyrighted material such as a book or music recording except that if the software is not copy protected you may a make one copy of...

Page 29: ...A Messtechnik GmbH makes no representations or warranties with regard to the Software and documentation including but not limited to the implied fitness for a particular purpose SITA Messtechnik GmbH...

Page 30: ...30 SITA Messtechnik GmbH Rev 5 engl 09 2012...

Reviews: