179
Irinos IR Users Manual © 2015 - 2016 Messtechnik Sachs GmbH, Germany
Troubleshooting and "first aid"
Stored in diagnostic
memory
Yes, can be disabled.
Event 27: Firmware update error
Description
An error occurred during the
firmware update process.
Type
Error
Reason / Solution
o
Invalid firmware file.
-> Use a valid firmware file.
o
Transmission error
-> Repeat the firmware update.
Note
If a firmware update fails, the old
firmware version remains active.
Reported to the user /
application
Yes, can be disabled.
Stored in diagnostic
memory
Yes, cannot be disabled.
Event 28: Firmware update successful
Description
Firmware successfully updated
Type
Information
Reported to the user /
application
No, cannot be enabled.
Stored in diagnostic memory
Yes, cannot be disabled.
Summary of Contents for Irinos IR
Page 2: ......
Page 7: ...Introduction...
Page 15: ...Safety instructions...
Page 22: ......
Page 23: ...System overview...
Page 39: ...Product descriptions...
Page 64: ......
Page 65: ...Pin assignment...
Page 92: ......
Page 93: ...Mounting Cabling...
Page 109: ...Setup...
Page 118: ...118 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Setup...
Page 120: ......
Page 121: ...Measurement Control via MscDll...
Page 168: ......
Page 169: ...Troubleshooting and first aid...
Page 187: ...Maintenance cleaning and disposal...
Page 190: ......
Page 191: ...Application notes...
Page 198: ......
Page 199: ...Specifications dimensions...
Page 209: ...209 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 210: ...210 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 211: ...211 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 212: ...212 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...