8.3.1
Current Loop
In the “Current Loop“ menu the range of the current output and the behavior of the current loop
during calibration and in case of an error are defined. Press <RIGHT> to scroll through the list
of the different modes. Select “OK“ and confirm with <ENTER> to confirm the selection of the
mode.
Current loop output modes:
• “I/O Mode,4-20mA“ The loop current ranges from 4mA to 20mA. 4mA correspond to the
starting point of the measuring range as defined in “IoutMin“, 20mA correspond to the
starting point of the measuring range as defined in “IoutMax“.
• “I/O Mode, 4-20mA (Err)“ This mode works in the analogous manner as the previously
described mode. In case of an error the current output is set to the value specified in the
submenu “I-Out Error“.
• “I/O Mode, 0-20mA“ The loop current ranges from 0mA to 20mA. 0mA correspond to the
starting point of the measuring range as defined in “IoutMin“, 20mA correspond to the
starting point of the measuring range as defined in “IoutMax“.