The first point of the submenu allows to select the unit. Pressing <ENTER> the alternates the
displayed unit between ppm and Vol.The value of P76 (see Chapter 8.4.1 Parameter) fixes the
number of digits after the point between 1 and 4 when displaying Vol.All internal calculations are
done in ppm. Access of measuring values via the RS-232 interface will always give the numbers
in ppm with 1ppm resolution.
8.1.1
Measuring Gas Setup
In the second line in the “Instrument Setup“ indicates the currently measured binary gas mixture.
For devices with Multi-Gas-Mode (MGM) pressing <ENTER> changes to the next binary gas
mixture included in the device. Up to 16 binary gas mixtures can be included. For every binary
gas mixture the corresponding set of linearization parameters and a set of calibration values is