55
STANDARD COMPLIANCE
MESI mTABLET ABI is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the above listed models can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the above listed models as recommended below,
according to the maximum output power of the communication
equipment.
Rated maximum
output power of
transmitter W
150 kHz to 80
MHz
d = 1.2√P
80 MHz to
800 MHz
d = 1.2√P
0.01
0.12
0.12
0.1
0.38
0.38
1
1.2
10
3.8
800 MHz to 2.5
GHz
d = 2.3 √P
Separation distance according to frequency of
transmitter m
0.23
0.73
1.2
2.3
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be
determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 4 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.