ERRORS
ERRORS
8
ERRORS
E1:
No oscillations detected.
There were no
oscillations detected.
If the error is repeated,
there is a possibility of
critical ischemia. Check the
placement of the cuffs and
repeat the measurement.
Error
Description
Solution
E2:
An anomaly has been
detected. The patient may
have moved during the
measurement process.
An anomaly has been
detected.
Remind the patient to remain
still during measurement and
repeat the measurement.
E3:
Inflation insufficiency.
Check the placement of
the cuffs and repeat the
measurement.
The cuff was insufficiently
inflated.
Check the placement of
the cuffs and repeat the
measurement.
E4:
An error occurred during
heart rate calculation.
An error occurred during
heart rate calculation.
Repeat the measurement.
if the error is repeated, the
measured value is outside
the measurement range of
the device.
E5:
An error occurred during
ankle brachial pressure
index calculation.
An error occurred
during ankle brachial
calculation.
Repeat the measurement.
if the error is repeated, the
measured value is outside
the measurement range of
the device.
E6:
Inflation error.
An error occurred during
inflation of the cuff.
Check the placement of
the cuffs and repeat the
measurement.
E7:
Deflation insufficiency.
Check the placement of
the cuffs and repeat the
measurement.
The cuff was insufficiently
deflated.
The cuff is not attached or is
not attached correctly. Attach
the cuff correctly and repeat
the measurement.
E8:
An error occurred during
systolic pressure calculation.
An error occurred
during systolic pressure
calculation.
Repeat the measurement.
if the error is repeated, the
measured value is outside
the measurement range of
the device.
E9:
An error occurred
during diastolic pressure
calculation.
An error occurred
during diastolic pressure
calculation.
Repeat the measurement.
if the error is repeated, the
measured value is outside
the measurement range of
the device.
E11:
A large pressure fluctuation
was detected. The patient
may have moved.
A large pressure
fluctuation was
detected. The patient
may have moved.
The results may be incorrect.
Repeat the measurement.