HDM97Pocket Quick Reference

 

 

\\

  

3

www.ibpmedical.com

Access Connectors 

Depending on the configuration the appliance part consists of a BNC inlet for pH electrode connection, a male connector 
for pressure transducer inlet, a connector for the flow probe and a directly attached cable which leads to the conductivity / 
temperature probe or connector for the probe. The appliance part is isolated against the USB-Connector withstanding 4000V 
AC or 6000V DC for one minute minimum according to ISO EN 60601-1.

USB CONNECTOR / CHARGING THE BATTERY

The USB connector is mounted on the right side of the instrument. It is implemented as a 
Mini-USB-A receptacle. The current consumption of the HDM97Pocket (all versions) can be up 
to 500 mA while charging the battery. The USB connector can be used to charge the battery and 
communicate with the instrument at the same time. The period for charging an empty battery to 
100% is approximately 3 hours. 

 

ATTENTION: The USB connector is not part of the appliance part. 

HDM97BL

HDM97BH

HDM97BN

HDM97BO

Flow

Pressure

Conductivity
Temperature

Conductivity
Temperature

Pressure

Conductivity
Temperature

Conductivity
Temperature

Pressure

Pressure

pH

Flow

Conductivity
Temperature

Pressure

Flow

HDM97BP

HDM97BQ

HDM97BQ 

with Option C (conductivity probe connector)

Summary of Contents for IBP HDM97Pocket Series

Page 1: ...M97BO HDM97BP HDM97BQ This document describes the major functions of the HDM97Pocket instrument series The complete manual can be found at http www ibpmt com medical_e meter_documents shtm and on the...

Page 2: ...tic discharge on the connectors This can lead to substantial damage of the instrument Make sure to be completely discharged before touching the connectors or cables connected to the instrument Adjust...

Page 3: ...USB connector is mounted on the right side of the instrument It is implemented as a Mini USB A receptacle The current consumption of the HDM97Pocket all versions can be up to 500 mA while charging the...

Page 4: ...P BQ Target time adjustment for pressure holt test PHT TIMER All Enables disables the stopwatch function Flow BL BQ Enables disables the Flow channel pH BP enables disables the pH channel TEMP BN BO B...

Page 5: ...and save setting or to discard this setting FLOW OK Only availble with version BQ ON OFF FLOW enabled FLOW disabled OK ESC Then to apply and save setting or to discard this setting TEMP OK Only avail...

Page 6: ...alue Reset of both the min And max Value Return to measuring mode ESC F F Channel Select No function TEMP F Available in versions bn bo bp and bq if enabled in the basic settings Selection between uni...

Page 7: ...x Value ESC Return to measuring display Reset accumulated volume to 0 Timer If enabled in the Basic Settings OK OK OK Afterwards Afterwards Timer Timer Timer The cycle start stop and reset can start a...

Page 8: ...n pass off through the outlet For immersion usage the probe must be unscrewed from the flow through adapter To accomplish that do not turn the upper shank the probe may be damaged Only turn the nut as...

Reviews: