Mesalabs IBP HDM97Pocket Series Safety Instructions And Quick Start Manual Download Page 2

HDM97Pocket Quick Reference

 

 

\\

  

2

www.ibpmedical.com

Safety Instructions 

For Your and Your Patients Safety Read and Follow Carefully the Safety Instructions Below 

• Read the entire manual (not only this guide) carefully before using the HDM97Pocket.

• Keep the instrument away from unauthorized persons.

•  Only use the HDM97Pocket as reference instrument to measure Conductivity, Temperature, Pressure, Flow and pH depending 

to it’s equipped features.

• Never use the HDM97Pocket as a replacement of sensors of a medical device.

• Do not use the HDM97Pocket in conjunction with a medical device during a treatment.

• Only use the instrument in a dry environment and do not touch it with wet hands.

• Make sure that no liquids intrude inside the instrument or it’s inlets / outlets.

• Always use a clean protection filter for the pressure transducer inlet.

• Prevent every mechanical overstraining of electrical wires. Never buckle or pull the cable of the conductivity probe.

• If the acquired values seem to be not believable, make sure that the HDM97Pocket is not defective.

•  Prevent electrostatic discharge on the connectors. This can lead to substantial damage of the instrument. Make sure to be 

completely discharged before touching the connectors or cables connected to the instrument.

•  Adjust the meter only, if you are familiar with the consequences of the adjustment. Consider the hints for reference solutions 

in the user manual.

•  Potentials above 42V against earth ground are dangerous. This potentials can lead to electrical shock and therefore to 

health hazards. Make sure that none of the connectors has higher voltage than mentioned before.

•  The direct connection of earth potential to any connector of the instrument is not applicable for safety purposes and may 

substantially damage the instrument. Make sure that the medical device itself is properly connected to earth ground 
according to its instruction manual.

• Make sure that the instrument never heats over 60°C / 140°F. Prevent direct sunlight.

• Never sterilize the instrument using an autoclave. Danger of explosion of the battery!

• Never throw the meter into a fire. Danger of explosion of the battery!

• Never open the meter. There are no serviceable parts inside.

•  Never try to replace or repair the internal battery. Handled in the wrong way these part is dangerous. Extreme risk of fire  

and explosion! 

Summary of Contents for IBP HDM97Pocket Series

Page 1: ...M97BO HDM97BP HDM97BQ This document describes the major functions of the HDM97Pocket instrument series The complete manual can be found at http www ibpmt com medical_e meter_documents shtm and on the...

Page 2: ...tic discharge on the connectors This can lead to substantial damage of the instrument Make sure to be completely discharged before touching the connectors or cables connected to the instrument Adjust...

Page 3: ...USB connector is mounted on the right side of the instrument It is implemented as a Mini USB A receptacle The current consumption of the HDM97Pocket all versions can be up to 500 mA while charging the...

Page 4: ...P BQ Target time adjustment for pressure holt test PHT TIMER All Enables disables the stopwatch function Flow BL BQ Enables disables the Flow channel pH BP enables disables the pH channel TEMP BN BO B...

Page 5: ...and save setting or to discard this setting FLOW OK Only availble with version BQ ON OFF FLOW enabled FLOW disabled OK ESC Then to apply and save setting or to discard this setting TEMP OK Only avail...

Page 6: ...alue Reset of both the min And max Value Return to measuring mode ESC F F Channel Select No function TEMP F Available in versions bn bo bp and bq if enabled in the basic settings Selection between uni...

Page 7: ...x Value ESC Return to measuring display Reset accumulated volume to 0 Timer If enabled in the Basic Settings OK OK OK Afterwards Afterwards Timer Timer Timer The cycle start stop and reset can start a...

Page 8: ...n pass off through the outlet For immersion usage the probe must be unscrewed from the flow through adapter To accomplish that do not turn the upper shank the probe may be damaged Only turn the nut as...

Reviews: