Install the Drive Unit – GMP 1000 ATEX and GMP 2000 ATEX
Complete
1
Fit screws and tighten clockwise (view from below).
q
yes
q
no
2
Install the drive unit into the tank plate. Install locking screws into the two oval
holes in the tank plate.
q
yes
q
no
3
Turn the drive unit clockwise (view from below) to the stop position.
q
yes
q
no
4
Tighten the two screws clockwise (view from below).
q
yes
q
no
Install the Drive Unit – GMP 5000 ATEX, GMP 10000 ATEX, GMP
20000 ATEX
Complete
1
Install the drive unit into the tank plate. Align the four holes in the drive unit to
the threaded holes in the tank plate.
q
yes
q
no
2
Screw the four bolts into the tank plate and tighten.
q
yes
q
no
Before Installing the Electrical Components
Complete
1
Ensure all incoming power is equipped with an emergency stop and an on/off
switch.
q
yes
q
no
2
Shield motor cables.
q
yes
q
no
3
Ensure the drive unit is grounded.
q
yes
q
no
4
Ensure electrical cables are long enough to enable the drive unit to be
disassembled and removed from the vessel.
q
yes
q
no
Install the AC Motor
Complete
1
Connect the drive unit.
q
yes
q
no
2
Ensure that the supply voltage and overload protection are set. Ensure that the
electrical cover and conduit entries are in place. Ensure that the outer drive
head rotates clockwise (view from above).
q
yes
q
no
Install the Revolution Counter
Complete
1
Connect the wiring from the tachometer into the appropriate apparatus
according to the proof of intrinsic safety.
q
yes
q
no
20