Operation
Display message
8. Open
door
If the leakage rate is above the limit value, then the display (and the print-out) will also show "Test
unsuccessful". In this case, follow the instruction in Section 7.3.
3.3 Test
run
____________________________________________________________
In order to check the operation of the autoclave under realistic conditions, a test run should be carried out with
the "Universal Program, 134°C wrapped" and a relevant load. After loading the autoclave and selecting the
program with the "Program" button, sterilization is started by pushing the "Start/Stop“ button. If the program
runs correctly, the following message will appear on the display (see Section 4.6):
with the values for the maximum values for pressure and temperature. If the immediate printout option has
been selected for an external printer a record of the program run will be printed.
3.4 Installation
record
___________________________________________________
As documentation that the autoclave has been set-up properly, an installation record should be produced by
an authorised person and a copy sent to MELAG. This is important in the event that you wish to make claims
under warranty provisions.
3.5 Safety
instructions
__________________________________________________
When opening the door, particularly after interrupting the drying process, residual steam can escape from
the autoclave chamber.
After opening the door, do not touch any metal surfaces - these will be hot! Danger of burns. Always use
the tray lifter to remove trays, or wear suitable hand protectors when taking out other items
We recommend the installation of a water leak detector (see installation details Section 2).
If you intend to install a water purification unit from another manufacturer, then consult MELAG before
you do so.
The appliance is not suitable for sterilizing liquids.
Under current VDE-regulations, this appliances is not suited for use in areas where there are risks of
explosion.
The appliance must only be serviced and repaired by MELAG or by its authorised representatives
(specialist dealers or customer services) using only original parts and following service instructions.
Before opening the housing always disconnect from the mains power supply!
In order to ensure effective sterilization with the autoclave observe the instructions in this User Manual,
and in particular ensure that the loading of the autoclave is appropriate for the program selected.
This appliance is only intended for use outside the patient environment (radius of 1.5 metres around the
treatment location).
Program Start-Stop
Vacuumtest
Universal-progr.
run successfully
Program Start-Stop