2.3
Levelling the autoclave
_______________________________________________
In order to ensure that condensate can drain out of the autoclave (which is important if it is to operate properly)
the appliance must have be clearly higher at the front than at the back. The autoclave should first be installed
in a horizontal position (this should be checked with a spirit level at the chamber flange) and then the front feet
should be extended by giving them at least five (5) turns.
6
3
2
4
8
9
10
1
7
5
11
Fig. 3 Typical installation of the Vacuklav
30-B with storage container for distilled / demineralised water, with
a wall-mounted trap
1
Vacuklav
30-B rear view
2
Cooling water inlet pipe (pressure pipe) with EN1717
3
Cooling water outlet pipe DN16
4 Power
connection
Power con
5
Water stop cock with safety combination in accordance with DIN 1988 (MELAG Art.-No.: 37310)
6
Leak detector with stop valve and sensor (MELAG Art.-No.: 01056)
7
Wall - mounted trap (MELAG Art.-No.: 37410)
8
Wall outlet (NW 40)
9
Storage container for distilled or demineralised water (MELAG Art.-No.: 244)
10
Feed water intake for distilled or demineralised water with filter
11
Y-junction with non-return valve (included in (7))