3 Setup and installation
23
Position Description
Art. no.
contained in
10.1
2x Cu seal 13.5x20
ME32050
Installation set
10.2
2x QSS-E straight
ME38710
Installation set
10.3
2x wastewater adapter G1/4" internal thread
ME56930
Installation set
11
Wall outlet NW40
--
present on the
building side
12
Compressed air supply
--
present on the
building side
12.1
Coupling plug for compressed air on 6 mm hose
ME80230
Installation set
13
Water tap
--
present on the
building side
14
Mains connection
--
present on the
building side
15
MELAdem 47 reverse osmosis unit
ME01047
--
15.1
Water inflow hose
ME37220
ME01047
15.2
PUR hose (black) 6/4 mm
ME28820
ME01047
15.3
Filter for MELAdem
ME48240
ME01047
15.4
Pressure tank MELAdem 47 (with shut-off valve and hose)
ME57065
ME01047
16
Safety combination HD
ME82384
--
17
Type AA safety device for separation from wastewater disposal in
accordance with EN 1717
--
present on the
building side
18
Water stop (leak water detector with cut-off valve and sensor)
ME01056
--
The following additional material can be ordered:
19
External tap for demineralised water
ME91900
--
PUR hose (black) 6/4 mm (10 m)
ME28820
--
Hose PTFE (8/6 mm, 5 m, wastewater hose)
ME39180
--
Connection to wastewater system
NOTICE
The Careclave connection set must be used for connection to the wastewater system, as other waste
water fittings may be made of materials that are not resistant to dental oil.
Since the backflow flap is not resistant to dental oil, the water drainage hose may only be connected to the
lowest connection as shown in the following illustration. The backflow flap may only be installed at the points
described.
PLEASE NOTE
For safe operation of the device, two drain hoses must be installed between the device and the
siphon.
n
The siphon must not be more than 2.5 m away from the device.
n
The hoses must be installed with a continuous slope.
1.
Cut two suitable drain hoses from the supplied PTFE hose.
2.
Remove the backflow flap from the double hose nozzle (pos. 3) for connecting the wastewater connection.
3.
Connect the overflow funnel (pos. 1) and the wastewater connection (pos. 3) of the device with separate hoses via
the double hose nozzle to a double chamber siphon.
4.
Perform a drain test by pouring approx. 500 ml into the overflow funnel.
Ê
The drain funnel must be empty within 30 s.
Summary of Contents for Careclave 618