![MELAG Careclave 618 Technical Manual Download Page 7](http://html1.mh-extra.com/html/melag/careclave-618/careclave-618_technical-manual_1766793007.webp)
2 Installation requirements
7
Additional space requirement for the feed water supply
Additional space is required for a storage container or a water treatment unit. It is also necessary to guarantee free ac-
cess to the hoses and cables leading from from the device to the water treatment unit.
Space require-
ments
MELAdem 47
Storage container
Osmosis module
Storage tank
Width
42 cm
Ø 24 cm
21 cm
Height
47 cm
51 cm
38 cm
Depth
15 cm
--
23 cm
Space is required above the MELAdem 53 / MELAdem 53 C for clear access to the hose connections.
Dimensions
MELAdem 53
MELAdem 53 C
Diameter
24 cm
24 cm
Height of the unit incl.
connecting parts
57 cm
45 cm
Requirements for the installation of a device
If the installation of the device is mandatory, one of the following measures must be implemented:
1.
It must be possible to pull out the device for operation (pos. a).
2.
There must be an exhaust air shaft in the rear area of the installation space that discharges the warm air upwards
(pos. b).
3.
There must be an exhaust air shaft in the rear area of the installation space that discharges the warm air to the rear
(pos. c).
4-5 cm
a
b
c
Mains supply
Implement the following safety measures when dealing with the cable and power plug:
n
Never damage or alter the power plug or cable.
n
Never bend or twist the power cable.
n
Never remove the plug by pulling on the power cable. Always take a grip on the plug.
n
Never place any heavy objects on the power cable.
n
Never run the power cable over areas in which it could become trapped (e.g. doors or windows).
n
Never lead the cable along a source of heat.
n
Never use any nails, paper fasteners or similar objects to fix the cable.
n
Should the power plug or cable suffer damage, switch off the device. The power cable or plug should only be
replaced by authorised technicians.
Summary of Contents for Careclave 618