9617722
50 / 60
10
Maintenance and cleaning
WARNING
– Danger to life from electric shock
Contact with live electrical parts on the appliance may result in serious injury
or death.
• Work on the electrical system must be conducted by a qualified electrician
who complies with the electrotechnical rules.
• Disconnect the appliance from the power supply before working on the elec-
trical system. To do this, turn the local mains switch to
OFF
and ensure that
it cannot be switched back on again.
• After switching off the repair switch the mains connection terminals are live!
CAUTION
– Material damage due to water ingress
• Never clean the washer-disinfector, control cabinets and electronic compo-
nents with water hose, high pressure cleaner or steam cleaner.
10.1 Maintenance
Note
In order to guarantee regular and correct maintenance, MEIKO recom-
mends concluding a maintenance contract with MEIKO Service.
Regular maintenance is a prerequisite for the long-term reliable and safe opera-
tion of a washer-disinfector. Maintenance which is neglected or improperly car-
ried out increases the residual risk of unforeseen damage to property and per-
sons, for which no liability will then be assumed.
For Austria, maintenance is equivalent to the periodic safety inspection, accord-
ing to ÖMPG and ÖMPBV.
Maintenance must be carried out:
• When the message No. 3
Perform maintenance
is displayed.
• After every 5000 cleaning cycles (factory setting, can be adjusted if re-
quired).
• At lease once a year (see
Next maintenance
sticker on the washer-disinfec-
tor).
Maintenance includes the following:
• Electrical safety according to VDE 0701/0702 (test report optional in Ger-
many)