In order to lower the temperature of the care utensils, a recooling step is carried
out. In this process, the high temperature is reduced with cool, filtered air and
the care utensils are dried.
At the end of the programme, the wash chamber door is unlocked. The care
utensils can be removed and must be visually checked by the operator.
The disinfection capability complies with the capability required DIN ISO 15883
Part 3, ”Requirements for and Testing of Cleaning/disinfection appliances using
Thermal Disinfection for Containers for Human Excreta” which specifies a mini-
mum A
0
value of 60.
Temperature curve of wash chamber/care utensils
t(s)
= Time in seconds
T(°C) = Temperature in °C
1 Precleaning with cold water
2 Cleaning steps 1-3 with hot water
3
Heating
4
Thermal disinfection A
0
60
5
Recooling/drying