7)
Check the effect achieved when the collector lens focus knob is adjusted
and continue to make minor adjustments to the arc position and the mirror
position until the light is as even in the FOV as possible.
8)
Remove the plastic slide and place a known good biological fluorescence
sample on the stage and again check the effect achieved when the collector lens
knob is adjusted and continue to make minor adjustments to the arc position and
the mirror position until the FOV displays an even an image as possible.
9)
If a shadow is seen the FOV, you do not have the field iris centered AND
the field iris not open all-the-way.
3.0 Operation
Once the microscope has been setup in its working location with all of the
components correctly installed, it is ready for use.
Do not install any bulb in your instrument other than the ones designated by Meiji
Techno.
For the Epi-Fluorescence Lamphouse, the following part number is the ONLY
recommended replacement bulb:
BA005
Replacement Mercury Lamp HBO 100W/2