41
3phase:
A_volts,B_volts,C_volts,A_amps,B_amps,C_amps
A_watts,B_watts,C_watts,A_vars,B_vars,C_vars
ph_Va_ref,ph_Vb_ref,ph_Vc_ref
ph_Ia_ref,ph_Ib_ref,ph_Ic_ref
A_pwrfact,B_pwrfact,C_pwrfact,freq
timer: status,time_count,A_volts,A_amps,freq
hi speed capture: ( all modes ) one line per sample
sample#,phase i.d.( A,B,C or S ), chan1, chan2,phase (chan1 to
chan2),watts,vars
query time: sends time
qt;
query date: sends date
qd;
Calibration
Correction Factors: correction factor read:
cfr,m;
read one correction factor specified by m ( 0 <= m <=53 )
cfrm=1
is returned where 1 is the integer value of the correction factor, or if m is
out of the valid range ‘invalid parameter’ message is sent.
correction factor write:
cfw,m,n;
write to one specified correction factor specified by m ( as with cfr ) ‘n’ is the
‘AOK
!
is returned if change is accepted or ‘invalid parameter’ is returned if either
m or n is out of range or string is not complete or improperly delimited.
send table showing all correction factors:
cfa;
write bogus test correction factors:
cfb;
update fudge float array:
cfu;
Correction Factor x Power MultiMeter Indication = 100 % accuracy
Instrument Correction Factor = 1 + (Correction Factor / 1,000,000)
Correction Factor = (Instrument Correction Factor -1) * 1,000,000
Voltage Correction Factors:
Ranges 3.75V 7.5V 15.0V 30.0V 75.0V 150.0V 300.0V
750.0V
A
phase:
cf0 cf1 cf2 cf3 cf4 cf5 cf6 cf7
B
phase:
cf8 cf9 cf10 cf11 cf12 cf13 cf14 cf15
C
phase:
cf16 cf17 cf18 cf19 cf20 cf21 cf22 cf23
Current Correction Factors:
Ranges 0.2A 0.4A 0.8A 1.6A 3.0A 6.0A 12.0A 25.0A 50.0A 100.0A
A
phase:
cf24 cf25 cf26 cf27 cf28 cf29 cf30 cf31 cf32 cf33
B
phase:
cf34 cf35 cf36 cf37 cf38 cf39 cf40 cf41 cf42 cf43
C
phase:
cf44 cf45 cf46 cf47 cf48 cf49 cf50 cf51 cf52 cf53
POWER MULTIMETER
Calibration Procedures
To check or re-calibrate any range in the Power MultiMeter simply do the following: