39
Setup commands
Single Phase custom input selection read:
slpcustomr;
sends channel 1 and 2 input source selections ( chan1=0, chan2=6 ) in 0-8
integer format.
0=Van, 1=Vbn, 2=Vcn, 3=Vab, 4=Vbc, 5=Vca, 6=Ia, 7=Ib, 8=Ic.
See Single Phase, pg 15.
Single Phase custom input selection write:
slpcustomw,m,n;
Write channel 1 and channel 2 input source selections ( same format as read)
m=channel 1, n=channel 2. Note: Once input source settings are changed the
change is not recognized until single phase mode is re-entered ( i.e. ‘ml’
command is re-issued ).
Three Phase mode phase angle reference source selection read:
s3pphrefr;
sends ( phref = 0 ) in 0-5 integer format. 0=Va, 1=Vb, 2=Vc, 3=Ia, 4=Ib, 5=Ic
See Three Phase, pg 20.
Three Phase mode phase angle reference source selection
write:
s3pphrefw,1;
writes in same format as s3pphrefr.
Note
: once phase angle reference source setting is changed the change is not
recognized until 3 phase mode is re-entered ( i.e. ‘m3y’ or ‘m3d’ command is
re-issued ).
Hi-Speed capture mode:
sends status i.e., after start or before stop mode setting
shscbar;
changes status,
K
=0 for after start,
K
=1 for before stop
shscbaw,K;
See High Speed Capture, pg 25.
Sends voltage and current range in any mode:
sranger;
format “vrng=K, irng=L”.
K=voltage range ( 0-7 ):
0 =3.75, 1 =7.5, 2= 15.0, 3=30.0, 4= 75.0, 5= 150.0, 6=300.0, 7=750.0
L=current range ( 0-9 ):
0 =0.2, 1 =0.4, 2 =0.8, 3 =1.6, 4 =3.0, 5 =6.0, 6 =12.0, 7 =25.0, 8 =50.0, 9 =100.0
See High Speed Capture, pg 25.
Change voltage and current range:
srangew,K,L;
Applies to timer and hi-speed capture mode ONLY. All other measurement
modes are autoranging ) Once these ranges are selected they are applied only
on entry to timer mode.
Sends start and stop trigger sources for timer and hi-speed
capture modes
striggerr;
will work in any mode yet applies only to timer and hi-speed capture modes.
Summary of Contents for PMM-1