To Save data directly to the
NOV
32
ram, at a set time interval proceed as follows:
1.
It is highly recommended that the NOV ram be cleared of all data if possible, this
is to allow maximum use of the unit’s ram memory (28,672 BYTES).
Refer to
page 30 of the instruction manual for number of bytes each test uses.
2.
From the Main Menu press the
More
key, then select the SetUp Page1 function
key to bring up the System SetUp Menu. Press the
F2
function key once to
toggle the selection to Save data at designated time. Press the OK key.
3.
Select the measuring screen desired and make the following changes.
4. Press
the
Set Save Time function key. From the TIME DATA LOG SETUP
screen use the arrow keys
⇑
⇓
⇐
⇒
to move the cursor to the optional fields.
Here the user can
Increase
or
Decrease
the Start/Stop time and the time
interval between each reading.
5. Pressing
the
Default SetUp function key presets the Start Time to the current
time and date, sets the Stop Time to terminate in 3 days and a Save time interval
of 30 minutes between readings.
6.
Once the settings have been completed, press the Save function key to toggle
Disab to
Enab.
Press the OK function key to initialize the settings. The Power
MultiMeter is now ready to begin saving the readings to the NOV ram memory.
7.
To stop the recording to the NOV ram at any time, simply press the Esc or Menu
key.
8.
To view the saved data, return to the Main Menu and select the View Saved Data
function key.