Battery indicator
The battery symbol on the LCD display contains four pairs of
segments. The battery is monitored continuously when the
instrument is turned on. The charge remaining in the battery, is
indicated by segment pairs as follows:
Fully charged battery
50% charged battery
Empty battery
Tests cannot be started, and the battery
may fail at any time
Symbol flashes when there is not
enough charge for a test and the
instrument will turn itself off.
When mains power is present the
indicator shows the battery is being
charged by animating the segments of the bar graph.
A blinking full battery icon indicates that the battery is
prevented from charging due to the temperature being out of
the allowable charge temperature range, 0 ºC to 40 ºC, or that
the battery has failed.
On screen error reporting
Should an error be detected during the operation of the
MIT515, MIT525, MIT1025 or MIT1525 an error code is
reported preceded by ‘Err’ with the read handbook warning.
Error codes are given in the following table.
‘Err’ code Fault
2
Output voltage over limit
3
FIFO (memory) overflow
4
HV board mismatch with control board setup
5
Battery low error
6
Control board detected inter-board
communication failure
7
Test button stuck
8
Measurement board i2c failed
9
Measurement board detected inter-board
communication failure
10
Isolation power supply cut-out
11
Instrument attempted auto power off but failed
12
HV circuit control fault
If an error occurs do not attempt to repair the instrument.
Obtain a repair number from Megger Instruments Limited,
carefully pack in a suitable box and send the faulty instrument
to the nearest Megger Approved Service Centre, if possible
noting the error that was reported.