background image

 

13 

   If a high level of electrical noise is detected the symbol [

] will be displayed. This indicates that excessive 

noise was detected during the test. If this remains displayed at the end of the test it is recommended the test is 
repeated. 

 

5.1.1.3 High current test - (All instruments) 

Performs a very fast loop test at a nominal 4A test current.  
 
Description: 

A resistive load (59 

Ω

) is used to measure voltage drop and calculate the loop resistance. If the loop contains 

inductance, the inductive reactance will be largely ignored. The loading current duration varies from 10 
milliseconds to 640 milliseconds depending on the test condition. 
 
When testing in close proximity to the sourcing transformer, a system phase angle of 18

°

 will cause an additional 

error of 5%. 

 

5.1.1.4 High resolution test   -   (LTW425 only) 

Performs a sequence of many high current loop tests to calculate a loop test result to three decimal 
places. 
 
Description: 

Alternate voltage measurements on loaded and unloaded half cycles using a load resistor of approximately 59 

Ω

as with the High Current test. The number of test cycles will be adjusted to obtain the optimum performance.  
 
When testing in close proximity to the sourcing transformer, a system phase angle of 18

°

 will cause an additional 

error of 5%. 

 

ALL TESTS WILL OPERATE PHASE TO EARTH, PHASE TO NEUTRAL AND PHASE TO PHASE. 

NOTE: 

THE LTW315 IS NOT RATED FOR PHASE TO PHASE MEASUREMENTS ON 240 VOLTS TO EARTH SYSTEMS

  

 

5.2 Performing a loop test 

Note: Factory settings for the instrument are: 

AUTO START  

 

enabled 

Mode 2 (extended testing)   enabled

  

 

5.2.1 Non-tripping loop test [No Trip] 

The [ No Trip]  range is a 0,01 

Ω

 resolution, low test current earth loop resistance measurement.  

 

5.2.2 No-trip testing using the mains socket test lead: 

The instrument is supplied with AUTO START enabled, and in Test Mode 2, which will automatically extend a test 
if noise is detected. 

 

5.2.2.1 Phase to earth loop test 

1. Select the [No Trip] test range on the lower range knob, and [Z] on the upper range knob where present. 
2. Connect the RED plug of the mains socket test lead to the instrument. 
3. Connect the GREEN plug of the mains socket test lead to the instrument. 
4. Insert the plug into the mains outlet socket. 
5. Ensure the supply voltage is displayed. 

The loop test will automatically start and the [

] symbol will appear, and remain on, for the duration of  

the test. 

6. After a test period of 10 seconds the measured loop value is displayed and the [

] symbol will disappear.  

To repeat the loop test press the [TEST] key. 

 

The symbols [

] and [

] may appear during the test. Refer to section 5.1.1.2 (Symbols) for an explanation of 

their meaning. 

 

5.2.2.2 Phase to neutral loop test

 

Phase to Neutral testing can be made using the No-Trip test, as described above for the Phase to earth test. 
However a High Current test will not trip RCDs when making a P-N test and it is recommended that a High 
Current test is used for this measurement. 

 

Summary of Contents for LTW315

Page 1: ...M LTW300 Series Earth Loop testers User Guide...

Page 2: ...lace whilst conducting tests Test leads probes and crocodile clips must be in good order clean and with no broken or cracked insulation On test leads fitted with a outlet plug and flying leads never c...

Page 3: ...oop test 14 High current Phase to Neutral loop test 14 High current Phase to Phase loop test 15 High resolution loop test LTW425 only 15 Prospective fault current display PFC 15 PFC measurement 15 Pos...

Page 4: ...nings and instructions before attempting to use the instruments This user guide describes the operation and functions of the following LTW300 series of loop impedance testers LTW315 LTW325 LTW335 LTW4...

Page 5: ...d return it to Megger as soon as possible to help us reduce any delays in supporting you should the need arise Carton contents for the LTW315 LTW325 LTW335 and LTW425 1 x LTW series loop tester 1 x 2...

Page 6: ...sults Fuse Blown Refer to User manual Battery status Ranges Supply noise warning R1 R2 Function selected Zref function selected required before R1 R2 can be measured Loop test active Temperature warni...

Page 7: ...n Backlight On Off Test lead connections See section 2 5 Loop test function all except LTW315 Z MaxZ R1 R2 Zref Neck strap slots Range selector NO TRIP High Current HIGH Res LTW425 Volts Frequency Set...

Page 8: ...ntrols OK Confirms a test result delete request ESC Aborts a result save process UP DOWN arrows Scrolls through circuit and distribution board numbers STORE Stores last test result LAST As next NEXT S...

Page 9: ...ifting up front panel tab 1 2 Foldaway underneath instrument 2 3 and push into retaining slot 4 Single or 3 Phase connection Earth Neutral or 2nd phase connection Warning Read user guide Maximum 300V...

Page 10: ...ssist drying 4 General operating instructions 4 1 Backlight operation Pressing the BACKLIGHT button will toggle the backlight on and off The backlight will automatically switch on at the start and end...

Page 11: ...zardous in a live socket outlet 4 3 Test lead connection The supplied test leads should be connected to the appropriate sockets on the rear of the instrument marked L0 L2 and L1 Standard test probes a...

Page 12: ...the noise symbol is displayed at the end of a test the test should be repeated or the test mode changed to mode 2 refer to section 8 SETUP procedure Mode 2 AUTO Default setting In Mode 2 the LTW loop...

Page 13: ...5 IS NOT RATED FOR PHASE TO PHASE MEASUREMENTS ON 240 VOLTS TO EARTH SYSTEMS 5 2 Performing a loop test Note Factory settings for the instrument are AUTO START enabled Mode 2 extended testing enabled...

Page 14: ...fast loop test WARNING This test is designed for non RCD protected circuits Due to the high currents used the instrument adds a brief pause between tests to prevent overheating This is indicated by th...

Page 15: ...nt display PFC NOTES The PSCC of a circuit is the largest Prospective Fault Current PFC In a single phase system this would be the larger of the earth loop PFC and the neutral loop PSCC In a multi pha...

Page 16: ...ortion of reactance in the measurement Worst case operating error calculated for Non Trip measurement mode in accordance with IEC 61557 3 1997 Intrinsic error or influence quantity Reference condition...

Page 17: ...ted 6 All subsequent loop test measurements will have the Zref resistance subtracted 7 To view the stored value of Zref press the PFC key press the PFC again to return to the R1 R2 value 8 If measured...

Page 18: ...Record number Each test result is assigned a unique record number from 0 to 999 logged automatically This cannot be changed by the user 7 2 To store a result 1 Make a Z MaxZ Zref or R1 R2 loop measure...

Page 19: ...nload from the software tool bar 4 Connect the tester to the PC using the USB download lead 5 Set the lower range knob to SEND 6 The instrument will automatically download its contents to the PC A bar...

Page 20: ...the function 4 Press the key to cycle through the options available for that function Once the desired option is displayed press TEST to store setting The display will show Ok and then return to the...

Page 21: ...Press the TEST button to store the offset The display will show Ok to confirm setting has been saved 15 The display will show the as a reminder to warn of there is a loop resistance offset selected To...

Page 22: ...ing test Retest or investigate source of noise Instrument needs calibration Further testing is possible but not recommended Noise symbol displayed at end of test indicates result may have been adverse...

Page 23: ...possible until batteries have been replaced Note NiMH rechargeable batteries show a lower charge than Alkaline batteries and may not give much warning before becoming exhausted Faulty stuck key or sw...

Page 24: ...n unused for an extended period 10 3 Fuse Blown indication If the fuse has blown after contact with a live supply the fuse blown symbol and the UnS message will be displayed and the instrument will st...

Page 25: ...0 03 100 V a c and 300 V a c LTW325 and LTW425 10 0 02 LTW425 15 0 03 50 V a c LTW325 LTW335 and LTW425 Nominal test current No Trip loop test 15 mA at nominal 230 V a c supply High current loop test...

Page 26: ...rdance with IEC61326 1 Operational uncertainties Refer to www megger com Power supply Battery 8 x 1 5 V cells IEC LR6 type AA alkaline Rechargeable 8 x 1 2V NiMH cells Battery life 2000 consecutive te...

Page 27: ...ted by a worldwide network of approved repair and calibration companies which offer excellent in service care for your Megger products Returning your product to Megger UK and USA service centres 1 Whe...

Page 28: ...com Megger Pty Limited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough Ont...

Reviews: