SAFETY WARNINGS
•
Safety Warnings and Precautions must be read and understood before
the instrument is used. They must be observed during use.
•
The circuit under test must be de-energized and isolated before
connections are made except for voltage measurement.
•
Circuit connections must not be touched during a test.
•
After insulation tests, capacitive circuits must be allowed to discharge
before disconnecting the test leads.
•
The Live Circuit Warning and Automatic Discharge are additional safety
features and should not be regarded as a substitute for normal safe
working practice.
•
Replacement fuses must be of the correct type and rating.
•
Test leads, including crocodile clips, must be in good order, clean and
have no broken or cracked insulation.
•
U.K. Safety Authorities recommend the use of fused test leads when
measuring voltage on high energy systems.
NOTE
THIS INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS
.