Capacitance testing is also conducted in a similar manner; though cable connections will need to
be adjusted during that process.
Lead cautions still apply, but you must modify connections of the RLC test leads to a
capacitor test configuration.
Preparing For Testing
Figure 14. Capacitance test lead connections.
4. Touch the
RLC test selection
icon then touch the
Capacitance test
icon in the popup menu.
5. Push and release the
Start (PTT)
button to run the capacitance test.
Baker DX Series Static Motor Analyzer
www.megger.com
12
1. Connect RLC test lead No. 1 to any motor lead.
2. Connect RLC test lead No. 2 to ground.
3. Disconnect RLC test lead No. 3 and
leave disconnected.