Figure 12. Resistance test start screen.
Preparing For Testing
Baker DX Series Static Motor Analyzer
www.megger.com
10
Resistance test procedures
1. Select the folder and record to use and check the
Status Bar
to ensure that the
Active Folder
and
Active Record
display the desired folder and record to be used.
2. Touch the
RLC Test Mode
icon.
3. The
Resistance Test
icon appears as the default RLC test selection.
4. If performing temperature compensation, touch the icon for the proper type of conductor
metal: aluminum or copper (
Copper
icon shown above).
5. If you need to define temperature compensation, touch the
Temperature
icon, then use the
popup keypad to input actual winding (not ambient) temperatures. Otherwise, skip to step 7.
6. The default temperature is 25° C; key in the actual winding temperature to use for
compensation and touch
Done.
NOTE
The default temperature to correct to is defined in the User Settings. Refer
to the Baker DX User Manual for more information.