80
3.
With the help of your healthcare provider, decide on the amount of insulin
(70 U-200 U) you need to insert into the Reservoir Patch.
4.
Draw air into the syringe up to the amount of insulin you need.
5.
Insert the needle into the insulin vial and push down on the plunger to
pressurize the vial.
6.
While still holding the plunger rod, flip the vial over so the vial is on top,
and then slowly pull down on the plunger to fill the syringe with the
number of desired units. Gently tap the side of the syringe to make air
bubbles rise to the top of the syringe. Slowly push up on the plunger just
enough to remove air bubbles from the syringe.
Warning:
Avoid using insulin from more than one vial because this may
introduce air into the syringe.
7.
With the vial down, hold the syringe. Pull straight up to remove the syringe
needle from the vial, and then insert it perpendicularly into the insulin fill
port on the side of the Reservoir Patch.
Warning:
Make sure that you insert the syringe perpendicularly rather than at
an angle into the insulin fill port.
8.
Keep the syringe vertical to the Patch Pump, and the needle inside the fill
port, pull back on the plunger until it is fully retracted. This will remove any
residual air from the reservoir. Bubbles will rise toward the plunger.