119
3.
Continue to set ISF segments as recommended by your healthcare provider.
Note:
Select segment, slide to left on each and a delete button will appear. Tap
delete
to delete the selected segment. The first segment cannot be deleted and
the starting time of first segment cannot be changed.
4.
When finished, press to save settings. Or Press to cancel setup and
exit the edit mode.
BG Target
A BG Target is your personal goal for keeping your BG levels under control. A BG
Target may be set as an actual range (with a low limit and high limit), or a single
value. Because your targets may vary throughout the day, you can set up to
eight BG targets. If you want to set just one target value instead of a range, set
both the low and high values to the same number.
If your current BG is above the BG Target Range, the bolus calculator feature
will calculate a correction dose. If your current BG is below the BG Target Range,
the bolus calculator will calculate a negative correction and thus subtract it from
your food bolus.
On the
Bolus Calc Setup
screen, select
BG Target
.
1
.
The first segment always starts at midnight. The
BG Range
field is indicated
by the blue editable digit.
Note:
You can set the low and high limit between 3.3 mmol/L and 13.9 mmol/L
(60 mg/dL and 250 mg/dL) with an increment of 0.1 mmol/L (1 mg/dL). The high
limit should never be lower than the low limit.