74
Chapter 4
Service C60/61
Error Message: C60/SERVICE (Transmitter Battery Low/End of Battery Life)
Cause: The Transmitter battery is low when C60 appears. The C60 service alarm will repeat each day
around 12:00 noon. Service C61 means that the Transmitter battery is used up.
Action: Press
the
SEL button and then the ACT button to clear the alarm. The battery cannot be changed in the
Transmitter. The battery life is about one year. Change out the Transmitter as soon as possible. Take off
the Sensor, and insert a new Sensor to connect to the new Transmitter. After the Transmitter is changed,
the Monitor must be reprogrammed with the new Transmitter ID number. See Chapter 2: Getting
Started.
Re Init
Error Message: C62/RE INIT (Re-Initialize Sensor)
Cause: The Monitor did not get the needed setup information from the Transmitter.
Action: Press
the SEL button and then the ACT button to clear the alarm. Disconnect the Sensor from the
Transmitter. Perform Search, reconnect Sensor, and following Initialization, enter blood glucose
calibration. See Chapter 2: Getting Started.
NOTE -
If this alarm is triggered repeatedly, change out Sensor and do Initialization process once more.