12 —1
12 — Printing and Data Ports
Description
The optional remote display is equipped with a thermal dot-matrix
printer which is located on the top of the unit. The Veris remote
display is capable of printing all vital signs parameters in tabular (text)
format. The remote display can also produce graphical (waveform)
formats as specified by the user.
NOTE: The main monitor does not have an internal printer. Printing
can only be done if the system has a remote display.
Snapshot Size
The period of time (in a graphical print) that starts 4.5 seconds prior
to pressing the PRINT key and lasting the duration of the Snapshot
setting in the print menu (6, 12, 18, or 24 seconds). (i.e. if the print
type is set to graphical and snapshot is set to 6 seconds then the
Veris prints out the waveform data selected 4.5 seconds prior to a
PRINT keypress and 1.5 seconds after the PRINT keypress).
History Size
History Size is a period of time defined by the user that is prior to the
snapshot time period. History Size can be selected as 6 or 12
seconds.
Safety
• Do not take the remote display into the MR Scanner room. The
remote display contains ferromagnetic material and can be
strongly attracted to the magnet causing a safety hazard.
NOTE: The remote display can only operate if the main monitor is
turned on. All audible and visual information displayed on the main
monitor is heard and seen on the remote unit in virtually real time. A
symbol displays in the upper right-hand corner of the screen to
indicate both devices are connected and active. The remote display
has access to all the same functions as the main monitor except for
the NIBP TEST and OPTION CONFIGURATION modes.
WARNING
!
!