Battery Specifications
Main Monitor Batteries
NOTE: Batteries in the main monitor are not externally accesible.
Replacement of batteries must be done by a qualified service
technician and the unit must be tested afterwards. Refer to the
Service Manual for battery replacement.
Number of Batteries:
2
Type:
Sealed lead acid
Voltage:
12 Volt
Amp-hour:
17 amp-hour
Battery Life:
Minimum of 10 hours on a fully charged battery.
Recharge time:
Approximately 12 hours, from a discharged voltage of 10.6
volts to a charged voltage of 13.7 volts at ambient
conditions (23°C ± 2°C).
Storage Time,
Shelf Life:
Charged battery, maximum 12 months with periodic
recharge at intervals specified by the manufactur