Vertical Maddox Rod Test
-
Purpose: Subjective test, to the Vertical Heterophoria in the way of
Maddox Rod.
-
Chart: Maddox Chart
-
Auxiliary Lens: Rotating Prism for right eye, Vertical Maddox for left
eye.
-
Objective: The horizontal bar seen with the left eye and the Maddox
Chart seen with the right eye must be united.
-
Operation Sequence
1.
Tap Maddox Chart
to choose the Horizontal
Maddox Rod Test Mode
2.
At this time, tap [Vertical Prism] button to change it into Vertical
Maddox Rod Test Mode.
3.
Turn the dial until the line and the dot are united as one.
- If the line locates over the dot, turn the dial clockwise (-
direction.) (Hypophoria)
- If the line locates under the dot, turn the dial counterclockwise
(+ direction.) (Hyperphoria)
- If the line and the dot are seen as united, that is an
orthophoria.