PCF Pro User Manual
DOC01851
Page 23 of 74
Installation Requirements
Machine Position
Site
Level floor and at least 150mm (6in) away from surrounding walls for ventilation (see
section 2-9 for dimensions and clearances).
There should be enough free space provided near the machine for safe and efficient
maintenance in the future.
Ambient Lighting
500 Lux minimum room lighting
Mains Supply
Cables shall be rated for maximum current of the equipment, meeting the requirements of IEC 60227 or IEC
60245. The PCF Pro is supplied with 2m of screened power cable.
200 - 230Vac 60Hz Models
(only)
Power cable which must be wired using a plug and socket, conductors colour coded as follows:
•
Live 1 Black
•
Live 2 Black
•
Live 3 Black
•
Earth Green/Yellow
For all Other Models
Power cable must be permanently installed by wiring directly into a switched wall panel. Under no
circumstances should it be wired using a plug and socket. Conductors colour coded as follows:
•
Live 1 Brown
•
Live 2 Black
•
Live 3 Grey
•
Neutral Blue
•
Earth Green/Yellow
The cable must be protected with a 3phase Residual Current Device (RCD) and Miniature Current Breaker
(MCB). The RCD and MCB shall only be used to supply this equipment.
The disconnect switch must:
•
Isolate all poles simultaneously
•
Be lockable in the off position
•
Comply with standards IEC/EN 61058-1 and IEC 60447 (creepage and clearances for 4kV transient
voltage)
•
Be situated close to the equipment for ease of accessibility, and in a position where the operator can
turn it off in an emergency without leaning over the machine.
•
Marked as the disconnection device for this equipment
•
Disconnect device off-position should be clearly marked (O) and the on-position should be marked (l)
To avoid risk of electric shock, this equipment must only be connected by a fully qualified
electrician and to a mains supply with a protective earth.
.
Supply Voltage
See section 5-1
Position of supply
<1.8m from power cable exit. (
Rear lower left of unit
)
If >1.8m please advise
Electro Magnetic
Compatibility (EMC)
Medical electrical equipment needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in the
Appendix
Summary of Contents for PCF Pro
Page 1: ...DOC01851 PCF Pro User Manual...
Page 7: ...DOC01851 Page 7 of 74 1 Introduction 1 Introduction User Manual...
Page 19: ...DOC01851 Page 19 of 74 2 Installation 2 Installation User Manual...
Page 26: ...PCF Pro User Manual DOC01851 Page 26 of 74 Installation Drawings Rear View Showing Services...
Page 27: ...PCF Pro User Manual DOC01851 Page 27 of 74 Front View...
Page 29: ...PCF Pro User Manual DOC01851 Page 29 of 74 Left Side View Lid Open...
Page 31: ...DOC01851 Page 31 of 74 3 Operation 3 Operation User Manual...
Page 55: ...DOC01851 Page 55 of 74 4 Troubleshooting 4 Troubleshooting User Manual...
Page 59: ...DOC01851 Page 59 of 74 5 Technical 5 Technical User Manual...
Page 65: ...DOC01851 Page 65 of 74 6 Appendices 6 Appendices User Manual...