PCF Pro User Manual
DOC01851
Page 38 of 74
Push up the locking flags
ENSURE
the thumb ball latches in the down
position before moving the carrier lift
Luer pipes
The PCF Pro basket comes with luer pipe
connections. Luer pipes are connected to
these connections for cleaning of lumen
instruments.
In general, the luer pipes should be replaced
every 12 months. However, depending on use
of the machine, the luer pipes may require
replacement sooner than 12 months.
The user must inspect the luer pipes before
every use and replace if required.
Note: Replace all luer pipes at the same
time.
Caution
•
Do not allow Luer pipes to
protrude from the basket.
•
If any instruments became
detached from the
Medisafe connection tubing
during processing, they
should be reprocessed
ensuring that the Luer
connection is firmly seated.
•
Check the condition of
basket irrigation port Luer
hoses, ensure they are not
kinked or ruptured.
The luer pipe clips into the luer connector.
To remove the luer pipe, press the luer
connector clip.
Luer connector
Luer connector clip
Luer pipe
Summary of Contents for PCF Pro
Page 1: ...DOC01851 PCF Pro User Manual...
Page 7: ...DOC01851 Page 7 of 74 1 Introduction 1 Introduction User Manual...
Page 19: ...DOC01851 Page 19 of 74 2 Installation 2 Installation User Manual...
Page 26: ...PCF Pro User Manual DOC01851 Page 26 of 74 Installation Drawings Rear View Showing Services...
Page 27: ...PCF Pro User Manual DOC01851 Page 27 of 74 Front View...
Page 29: ...PCF Pro User Manual DOC01851 Page 29 of 74 Left Side View Lid Open...
Page 31: ...DOC01851 Page 31 of 74 3 Operation 3 Operation User Manual...
Page 55: ...DOC01851 Page 55 of 74 4 Troubleshooting 4 Troubleshooting User Manual...
Page 59: ...DOC01851 Page 59 of 74 5 Technical 5 Technical User Manual...
Page 65: ...DOC01851 Page 65 of 74 6 Appendices 6 Appendices User Manual...