11
User manual for medifa 5000 operating table Version 2.1
Safety instructions and user obligations
The medifa 5000 is suitable for use with HF surgical units, defibrillators and defibrillator monitors. Please adhere
to the user manual and safety instructions given by the manufacturers of these devices!
5.4
High-frequency (HF) surgical devices and defibrillators
HAZARD
Danger of burns for patients!
When using HF surgical unit, defibrillators and defibrillator monitors, there is a risk of burns for the patient if
no safety measures are taken.
Here are the safety measures to be followed:
•
Place the patient on medifa 5000 isolated from metal parts (medifa 5000, accessories) and conductive pad-
dings or tubes.
•
Ensure that the patient does not come into contact with wet cloths or underlays. Use only dry materials!
5.5
Maintenance and repair
To avoid failure faults and to ensure operational safety, annual maintenance must be carried out so that the
functioning is fully maintained. It is recommended to hire a medifa-certified service technician for this mainte-
nance procedure.
Repairs
Repairs may only be carried out by medifa's technical customer service department or by personnel authorised,
trained and certified by medifa.
medifa shall not be liable for any damage caused due to deferred inspections, faulty repair or maintenance, or
changes made to the product.
For service work, please contact the medifa technical customer service department.
5.6
Service life
The medifa 5000 has a service life of 10 years if the specified inspection is adhered to by the medifa service de-
partment. This excludes all wear parts and padding.