Preventive Maintenance Plan 

 

The TEMP 3

 temperature controller should be verified at least once every five years.  Use of 

the following items to accurately measure the bag temperature for verification. 

 

Fluke 51II Digital Thermometer   

 

6” Long .093” Diameter Type T Needle Thermocouple 

 

 

(with the Thermocouple calibrated to the Thermometer) 

With the outer wrap still on a 1000ml solution bag, pierce the bag through the needle injection 
port on the bag with an 18ga needle to make it easier to insert the Thermocouple.  Insert the 
Thermocouple through the pierced needle injection port.  Place the bag into the TEMP 3

 

through the lower door of the cabinet.  Feed the Thermocouple wire out of the cabinet to the 
Fluke Thermometer and position the probe in the center of the bag. Close the lower door.  Add 
five additional bags to the cabinet from the top loading door. 

Turn the power to the TEMP 3

 on.  Adjust the temperature controller set-point temperature to 

40

C.  Allow the unit to operate for a minimum of 1 hour before verifying the temperatures.  

After the allotted time, compare the temperature on the controller to the temperature reading on 
the Fluke Thermometer.  The two temperatures should be within one degree Celsius of each 
other.  If the two temperature readings are within one degree Celsius of each other the 
temperature controller has been verified.  If they are not, please contact Medical Solutions, Inc. 
at 800-435-7600.   

If a Fluke Digital Thermometer is not available a glass laboratory thermometer may be used as 
an alternative.  Load the unit with (6) 1000ml bags and turn the power on. Adjust the 
temperature controller set-point temperature to 40

C.  Allow the unit to operate for a minimum 

of 1 hour before verifying the temperatures.  Remove the bottom bag and place the end of the 
glass thermometer on top of and in the middle of the bag.  Fold the bag over the thermometer and 
hold for approximately two minutes or until the thermometer stops moving.  If the reading on the 
thermometer is between 38

C and 41

C the temperature controller has been verified. If it is not, 

please contact Medical Solutions, Inc., at 800-435-7600. 
 

 

Summary of Contents for TEMP 3 7000 Series

Page 1: ...5595 A TEMP 3 FLUID WARMER OPERATORS MANUAL 7000 Series Medical Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 435 7600 703 834 0039 Fax www warmiv com ...

Page 2: ...sed to regulate the heating of the fluid bags and the over temperature safety The top display number in red will indicate the temperature of the bag currently being warmed while the bottom display number in green displays the desired set point temperature During the time that the controller is directing the unit to warm a light will illuminate in the upper left corner of the display The Documentor...

Page 3: ... Ambient to 43 C Factory Preset Temperature 40 C Ambient Operating Conditions Storage Temperature 40 to 80 C 40 to 176 F Humidity Conditions 0 to 90 up to 40 C non condensing Operating Temperature 10 to 55 C 14 to 131 F Atmospheric Pressure during minimum of 16 288 in Hg Transporting Electrical Requirements Power requirements 110 120 VAC 60 HZ Fused 2 each 2 amp Amperage consumed 1 25 amps Heater ...

Page 4: ...mounting bracket 4 Mount the bracket securely in a convenient place on a wall or cart The TEMP 3 loaded with 6 bags of solution will weigh approximately 30 pounds When mounting to a wall care should be taken to secure the bracket using suitable bolts for the wall material to support a minimum of 90 lbs 5 Move the TEMP 3 to the desired location check that the power switch is off and plug the power ...

Page 5: ...cabinet 2 Turn the unit on by depressing the power switch The switch will illuminate when energized The controller will perform a self test and then it will display the current temperature of the bottom bag The factory temperature setting is 40 degrees C To change the temperature setting follow the instructions below 3 The controller will display 2 temperatures in its display The top display numbe...

Page 6: ... button until the number 8 appears in the fourth position When the display reads 8088 Tap the right button 10 When the display flashes all four digits Tap the right button 11 The display flashes P1 date 12 Tap the right button and the small digits on the left representing the year will flash 13 Tap the left button until the desired number appears and then Tap the right button The digits representi...

Page 7: ...o qualified service personnel Have a qualified person remove all liquid and dry the components Perform any necessary hospital electrical safety checks before returning the unit to service Trouble Shooting 1 If the unit fails to warm do the following a Check to see that the unit is plugged in and the power switch is on b Check the set point temperature setting so that the controller is directing th...

Page 8: ...w the unit to operate for a minimum of 1 hour before verifying the temperatures After the allotted time compare the temperature on the controller to the temperature reading on the Fluke Thermometer The two temperatures should be within one degree Celsius of each other If the two temperature readings are within one degree Celsius of each other the temperature controller has been verified If they ar...

Page 9: ...ar and partial side panel to access the electrical compartment Remove 3 screws 2 on the side and 1 on the back Remove 5 screws 1 on the back and 4 on the side Remove 3 screws on the bottom Service panel location and removal ...

Page 10: ...5595 A 10 Fuse removal instructions To replace the fuses disconnect the power cord from the unit Use a flat screwdriver to unlock the fuse holder from the Power Inlet Replace the fuses as marked ...

Page 11: ... CONSEQUENTIAL OR OTHER DAMAGES INCLUDING BUT NOT LIMITED TO LOST PROFITS To enable MSI to administer the Warranty properly Buyer shall 1 register the Warranty using the form provided below and 2 promptly notify MSI of any claim hereunder As a condition to the Warranty Buyer shall use and maintain the Equipment according to the provisions of the Operator s service manual IMPORTANT THIS IS YOUR TEM...

Reviews: