Cleaning Instructions
1.
Turn the unit off and disconnect it from the power source.
2.
Wipe down the cabinet with a damp cloth using a general hospital cleaner (Isopropyl
Alcohol) as needed, following the manufacturer’s instructions for the cleaner.
3.
Avoid contact with the electrical connections and the electrical components.
4.
In the event that excessive fluid enters the cabinet, disconnect power to the unit and refer to
qualified service personnel. Have a qualified person remove all liquid and dry the
components. Perform any necessary hospital electrical safety checks before returning the
unit to service.
Trouble Shooting
1.
If the unit fails to warm, do the following:
a.
Check to see that the unit is plugged in and the power switch is on.
b.
Check the set-point temperature setting so that the controller is directing the unit to
warm.
c.
Check to see that the minimum number of I.V. bags is loaded in the cabinet.
2.
If the unit overheats call Medical Solutions, Inc.
3.
Contact Medical Solutions, Inc., for any additional technical assistance @ 800-435-7600.