Medical Solutions TEMP 3 7000 Series Operator'S Manual Download Page 6

5595.A 

6

NOTES 

The TEMP 3

 may be left on continually without harm to the unit.  It is suggested that each 

facility should contact the fluid manufacturer to obtain the manufacturer’s specifications for time 
and temperature for the fluids being warmed.   

The TEMP 3

 is not intended for Blood or Blood Products.   

In the event that fluid should spill inside the cabinet – unplug the unit to prevent an 
electrical shock hazard and refer to step #4 of the Cleaning instructions. 

 

The Documentor

®

 Resetting Instructions 

TO CHANGE THE DATE: 

1.

 

Plug The Documentor

®

 into an electrical outlet. 

2.

 

Press and hold blue RESET button located on the bottom of the The Documentor

®

3.

 

While holding down the RESET button, press and hold the 

 right button on the front panel. 

4.

 

While holding the 

 right button, release the RESET button. 

5.

 

When “

Prog” 

appears in the display, release the 

 right button,

 Prog 

finishes. 

6.

 

Tap the 

 right button again.  The display will show “0000” and the 1

st

 digit will flash. 

7.

 

Tap the   left button until the number “8” appears, then Tap the 

 right button twice so the third 

“0” is flashing. 

8.

 

Tap the   left button until the number “8” appears in the third position then Tap the 

     

 right button.  

9.

 

Tap the 

 left button until the number “8” appears in the fourth position. When the display reads 

“8088”, Tap the 

 right button. 

10.

 

When the display flashes all four digits, Tap the 

 right button. 

11.

 

The display flashes “

P1(date)”

12.

 

Tap the 

 right button and the small digits on the left, representing the year, will flash. 

13.

 

Tap the   left button until the desired number appears and then Tap the 

 right button. The digits 

representing the month should flash. 

14.

 

Tap the   left button until the desired number appears and then Tap the 

 right button. The digits 

representing the day should flash. 

15.

 

Tap the   left button until the desired number appears and then Tap the 

 right button.  

16.

 

When the desired date is set, Tap the 

 right button until the display flashes “

P1 (date)”

17.

 

Tap the blue RESET button to return to the standard display. 

 

TO CHANGE THE TIME: 

18.

 

 Follow steps 1-11. 

19.

 

 Tap the   left button.  When the display flashes “

P2 cloc”

, Tap the 

 right button. 

20.

 

 The hour format will flash (12 or 24).  If necessary, tap the 

 left button until 24 flashes. Tap the 

 right button. 

21.

 

The hour digits will flash.  Tap the   left button until the desired setting is achieved. (the hour 

must be set in military time 3:05 pm = 15:05) Tap the 

 right button 

22.

 

 The minutes digits will flash.  Tap the 

 left button until the desired setting is achieved.  Tap the 

 right button twice. 

23.

 

 The display will flash “P2 Cloc”, Press and release the blue RESET button. 

Summary of Contents for TEMP 3 7000 Series

Page 1: ...5595 A TEMP 3 FLUID WARMER OPERATORS MANUAL 7000 Series Medical Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 435 7600 703 834 0039 Fax www warmiv com ...

Page 2: ...sed to regulate the heating of the fluid bags and the over temperature safety The top display number in red will indicate the temperature of the bag currently being warmed while the bottom display number in green displays the desired set point temperature During the time that the controller is directing the unit to warm a light will illuminate in the upper left corner of the display The Documentor...

Page 3: ... Ambient to 43 C Factory Preset Temperature 40 C Ambient Operating Conditions Storage Temperature 40 to 80 C 40 to 176 F Humidity Conditions 0 to 90 up to 40 C non condensing Operating Temperature 10 to 55 C 14 to 131 F Atmospheric Pressure during minimum of 16 288 in Hg Transporting Electrical Requirements Power requirements 110 120 VAC 60 HZ Fused 2 each 2 amp Amperage consumed 1 25 amps Heater ...

Page 4: ...mounting bracket 4 Mount the bracket securely in a convenient place on a wall or cart The TEMP 3 loaded with 6 bags of solution will weigh approximately 30 pounds When mounting to a wall care should be taken to secure the bracket using suitable bolts for the wall material to support a minimum of 90 lbs 5 Move the TEMP 3 to the desired location check that the power switch is off and plug the power ...

Page 5: ...cabinet 2 Turn the unit on by depressing the power switch The switch will illuminate when energized The controller will perform a self test and then it will display the current temperature of the bottom bag The factory temperature setting is 40 degrees C To change the temperature setting follow the instructions below 3 The controller will display 2 temperatures in its display The top display numbe...

Page 6: ... button until the number 8 appears in the fourth position When the display reads 8088 Tap the right button 10 When the display flashes all four digits Tap the right button 11 The display flashes P1 date 12 Tap the right button and the small digits on the left representing the year will flash 13 Tap the left button until the desired number appears and then Tap the right button The digits representi...

Page 7: ...o qualified service personnel Have a qualified person remove all liquid and dry the components Perform any necessary hospital electrical safety checks before returning the unit to service Trouble Shooting 1 If the unit fails to warm do the following a Check to see that the unit is plugged in and the power switch is on b Check the set point temperature setting so that the controller is directing th...

Page 8: ...w the unit to operate for a minimum of 1 hour before verifying the temperatures After the allotted time compare the temperature on the controller to the temperature reading on the Fluke Thermometer The two temperatures should be within one degree Celsius of each other If the two temperature readings are within one degree Celsius of each other the temperature controller has been verified If they ar...

Page 9: ...ar and partial side panel to access the electrical compartment Remove 3 screws 2 on the side and 1 on the back Remove 5 screws 1 on the back and 4 on the side Remove 3 screws on the bottom Service panel location and removal ...

Page 10: ...5595 A 10 Fuse removal instructions To replace the fuses disconnect the power cord from the unit Use a flat screwdriver to unlock the fuse holder from the Power Inlet Replace the fuses as marked ...

Page 11: ... CONSEQUENTIAL OR OTHER DAMAGES INCLUDING BUT NOT LIMITED TO LOST PROFITS To enable MSI to administer the Warranty properly Buyer shall 1 register the Warranty using the form provided below and 2 promptly notify MSI of any claim hereunder As a condition to the Warranty Buyer shall use and maintain the Equipment according to the provisions of the Operator s service manual IMPORTANT THIS IS YOUR TEM...

Reviews: