Read and understand the Operators Manual before using the TEMP 3

 Fluid Warmer. 

 
Intended Use

 

The TEMP 3

 Fluid Warmer is designed to store, rapidly warm, and maintain the 

temperature of I.V. and irrigation fluid bags prior to their use.   

The TEMP 3

 is designed to supply warmed I.V. and irrigation fluid bags for a single 

room, (i.e., Operating, LDR, Patient's, etc.).  It is not intended to supply warmed I.V. 
bags or irrigation solution bags to more than one room at a time. 

 

It is suggested that each facility should contact the fluid manufacturer to obtain the 

 

manufacturer’s specifications for time and temperature for the fluids being 

 warmed. 

Unit Description   

 

The TEMP 3

 holds up to (6) 1000ml or (6) 500 ml bags and can be attached to the wall, 

a K740 stand, or an anesthesia cart.  A temperature controller is used to regulate the 
heating of the fluid bags and the over-temperature safety. The top display number (

in 

red

) will indicate the temperature of the bag currently being warmed while the bottom 

display number (

in green

) displays the desired set-point temperature. During the time 

that the controller is directing the unit to warm, a light will illuminate in the upper left 
corner of the display.   

The Documentor

®

 is designed to assist in the compliance with JCAHO

 standard MM 

2.20 and the I.V. bag manufacturer’s specifications for time and temperature. 

 

Important Labels and Warnings: 

Model/Serial Number label 

Caution:  Replace fuses as marked. 

Caution:  Electric shock hazard, do not remove protective panel.  Refer servicing to Qualified  
 

    Personnel. 

Caution:  Verify temperature of warmed items before use. 

Danger:   Risk of explosion if used in the presence of flammable anesthetics. 

Warning: Unit not intended for blood or blood products. 

Caution:  Minimum number of bags required for proper operation: 
   500ml 

 

 

          1000ml - 2 

Caution: Use approved cord only. 

 

 

Summary of Contents for TEMP 3 7000 Series

Page 1: ...5595 A TEMP 3 FLUID WARMER OPERATORS MANUAL 7000 Series Medical Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 435 7600 703 834 0039 Fax www warmiv com ...

Page 2: ...sed to regulate the heating of the fluid bags and the over temperature safety The top display number in red will indicate the temperature of the bag currently being warmed while the bottom display number in green displays the desired set point temperature During the time that the controller is directing the unit to warm a light will illuminate in the upper left corner of the display The Documentor...

Page 3: ... Ambient to 43 C Factory Preset Temperature 40 C Ambient Operating Conditions Storage Temperature 40 to 80 C 40 to 176 F Humidity Conditions 0 to 90 up to 40 C non condensing Operating Temperature 10 to 55 C 14 to 131 F Atmospheric Pressure during minimum of 16 288 in Hg Transporting Electrical Requirements Power requirements 110 120 VAC 60 HZ Fused 2 each 2 amp Amperage consumed 1 25 amps Heater ...

Page 4: ...mounting bracket 4 Mount the bracket securely in a convenient place on a wall or cart The TEMP 3 loaded with 6 bags of solution will weigh approximately 30 pounds When mounting to a wall care should be taken to secure the bracket using suitable bolts for the wall material to support a minimum of 90 lbs 5 Move the TEMP 3 to the desired location check that the power switch is off and plug the power ...

Page 5: ...cabinet 2 Turn the unit on by depressing the power switch The switch will illuminate when energized The controller will perform a self test and then it will display the current temperature of the bottom bag The factory temperature setting is 40 degrees C To change the temperature setting follow the instructions below 3 The controller will display 2 temperatures in its display The top display numbe...

Page 6: ... button until the number 8 appears in the fourth position When the display reads 8088 Tap the right button 10 When the display flashes all four digits Tap the right button 11 The display flashes P1 date 12 Tap the right button and the small digits on the left representing the year will flash 13 Tap the left button until the desired number appears and then Tap the right button The digits representi...

Page 7: ...o qualified service personnel Have a qualified person remove all liquid and dry the components Perform any necessary hospital electrical safety checks before returning the unit to service Trouble Shooting 1 If the unit fails to warm do the following a Check to see that the unit is plugged in and the power switch is on b Check the set point temperature setting so that the controller is directing th...

Page 8: ...w the unit to operate for a minimum of 1 hour before verifying the temperatures After the allotted time compare the temperature on the controller to the temperature reading on the Fluke Thermometer The two temperatures should be within one degree Celsius of each other If the two temperature readings are within one degree Celsius of each other the temperature controller has been verified If they ar...

Page 9: ...ar and partial side panel to access the electrical compartment Remove 3 screws 2 on the side and 1 on the back Remove 5 screws 1 on the back and 4 on the side Remove 3 screws on the bottom Service panel location and removal ...

Page 10: ...5595 A 10 Fuse removal instructions To replace the fuses disconnect the power cord from the unit Use a flat screwdriver to unlock the fuse holder from the Power Inlet Replace the fuses as marked ...

Page 11: ... CONSEQUENTIAL OR OTHER DAMAGES INCLUDING BUT NOT LIMITED TO LOST PROFITS To enable MSI to administer the Warranty properly Buyer shall 1 register the Warranty using the form provided below and 2 promptly notify MSI of any claim hereunder As a condition to the Warranty Buyer shall use and maintain the Equipment according to the provisions of the Operator s service manual IMPORTANT THIS IS YOUR TEM...

Reviews: