5595.A
5
The Documentor
®
Instructions
Prior to loading the cabinet with I.V. bags, use The Documentor
®
to label the I.V. bags with a
future date that follows the manufacturer’s recommendation for the length of time a bag is to
be warmed.
“Future Date Label”
To create a future date label (See The Documentor
®
Resetting Instructions for adjusting the clock
date):
i)
Hold the Label sheet face up;
ii)
Line up the arrow on The Documentor
®
under the words “Remove Fluid By:” on the
label;
iii)
Push the label sheet towards the arrow until The Documentor
®
is activated; and
iv)
Place a label on each I.V. bag for documentation and compliance purposes.
Load the labeled bags of solution into the TEMP 3
through the top door.
Operating Instructions
1.
Load the labeled bags of solution into the cabinet through the top of the cabinet.
2.
Turn the unit “on” by depressing the power switch. The switch will illuminate when
energized. The controller will perform a self-test and then it will display the current
temperature of the bottom bag. The factory temperature setting is 40 degrees C. To change
the temperature setting, follow the instructions below.
3.
The controller will display 2 temperatures in its display. The top display number (in red)
will indicate the current temperature of the bottom bag in the cabinet. The bottom display
number (in green) will indicate the set–point temperature (40
C). During the time that the
controller is directing the unit to warm a light will illuminate in the upper left corner of the
display.
4.
To change the temperature setting, follow the instructions below.
TO CHANGE THE SET POINT
i)
Press the index key (left most key) to view set-point one (sp1);
ii)
press the up or down arrow keys to change the set point;
iii)
press the enter key (right most key) to accept the new set point; and
iv)
press the index key