38
Cleaning and disinfection
CAUTION
Invia Liberty pump with associated products (docking station, rail holder and charger) should be
cleaned / disinfected after every use. Before cleaning the device, unplug the pump from the wall
outlet.
CAUTION
Invia Liberty pump, docking station, rail holder, charger and carrying case cannot be sterilized.
Immersion disinfection, thermal disinfection and ultrasound cleaning are not permitted.
Disinfection (Pump housing, docking station and charger)
CAUTION
Invia Liberty pump can be disinfected with “alcohol”.
Disinfection
Invia Liberty can be disinfected with the disinfecting agent group “alcohol”.
Do not use other cleaning agents (e.g. Terralin) as they can damage the plastic housing.
Immersion disinfection, thermal disinfection and ultrasound cleaning are not permitted.
Sterilization
Invia Liberty and Invia Liberty accessories cannot be sterilized.
CAUTION
Do not use other cleaning agents (e.g. Terralin) as they can damage the plastic material.
Cleaning
Disinfection
Sterilization
Dishwasher
Pump
Docking station
Rail holder
Charger
Carrying case*
P
P
O
O
Wipe off with
a damp cloth.
Wipe off with a dis-
infecting agent
Sterilization / cleaning in a dishwasher is
NOT allowed.
* If needed, the carrying case can be washed in washing machine
Cleaning procedure for Invia Liberty pump, docking station and charger
1.
Wear suitable protection (clothing, gloves, face mask and goggles) according to local guidelines.
2.
Apply the disinfectant agent in accordance with instructions from the manufacturer
1)
.
Pay attention particularly to edges, narrow corners and bottom side.
3.
Leave the disinfectant on. Follow the recommended residence time for the disinfectant
as instructed by the manufacturer
1)
.
4.
Thoroughly clean the surface, all edges, housing niches, corners, bed holder, brackets,
port covers and bottom side.