Fault number
Problem description
on the display
Troubleshooting
on the display
Pressure
Remarks / potential cause of fault
W
a
rnin
g
401
P
Recharge the battery either by placing the Invia Liberty pump in the
docking station or plug in the charger to the electrical outlet port on the pump.
Remaining time of battery is approximately 30 minutes.
402
Unplug USB cable
405
O
If the pump is in Standby mode for more than 5 minutes, an alarm will go off.
To continue therapy press
“On”
[
] or switch off the pump by pressing [
> 3 seconds ].
406
P
Cool Invia Liberty pump down as per instructions on the display.
A
la
rm
301
P
Dressing:
- Check dressing for air leakage. Press firmly around the edges of the dressing, around
the drain tubing or on the Invia FitPad.
- Apply some additional film dressing to seal the leaking area.
Connectors:
- Ensure that the dressing tubing is connected properly to the canister tubing.
- Ensure that the canister tubing is inserted straight into the pump.
Canister:
- Ensure that the canister is properly inserted, release the canister and reposition.
- Ensure that the orange O-ring / gasket, placed beside the canister tubing on the pump
is not missing. Additional O-ring is available via Medela customer service.
302
P
Tubing:
- Ensure that the tubing is not twisted, kinked or clamped.
- If the canister tubing is clogged, change the tubing.
Canister:
- If canister is full or filter clogged, replace canister.
- If dressing tubing is clogged, change the dressing.
Then press "On" to restart the therapy.
305
O
Recharge the battery either by placing the Invia Liberty pump in the docking station or plug in the
charger to the electrical outlet port on the pump.
Remaining time of battery is 15 minutes.
306
P
Change the canister, see chapter “Change Invia Liberty canister and Invia Liberty tubing”.