35
EN
O
P
OPERATION HALTS
OPERATION CONTINUES
Fault number
Problem description
on the display
Troubleshooting
on the display
Pressure
Remarks / potential cause of fault
A
la
rm
311
O
Canister:
- Release the canister and reposition.
Tubing:
- Ensure that the tubing is not twisted, kinked or clamped.
312
O
Occurs when alarm 301 was pending for 5 minutes.
For troubleshooting, refer to instructions for Alarm 301.
Restart the pump by pressing
“On”
[
].
313
O
To restart the therapy after canister has been changed, press right selection button
“On”
[
].
315
O
Occurs when Warning 406 was pending for 30 minutes.
Cool Invia Liberty pump down.
Int
e
rn
a
l f
a
ul
t
O
Restart the pump. If internal fault remains, turn off by pressing [
> 3 seconds ] and
contact the Medela customer service.