Multitest 0.5dV, 1dV & 2.5dV
Service and Repair Manual
Number 022-780
Issue Number: 02
Page 26 of 32
17. Fault messages
84.
“Software Timeout” – The
software has previously
crashed
85.
“
Software Timeout
Over
travel detected
” – The
hard limit switch has been
hit
86.
“
Motor 0: overtemp
” – The
temperature sensor has
detected that the motor is
over temperature.
Over travelled / Emergency stop
Faulty Emergency Stop Loom
.
The emergency stop has two parts a hardware stop circuit and software stop circuit.
The hardware stop circuit and overtravel circuit both use the same signal to trigger a
relay to stop the stand. So if the hardware wire of the Emergency stop loom is
damaged/faulty it will activate the same relay. The software detects this and then
checks to see if it has received a software emergency stop signal.