Multitest 0.5dV, 1dV & 2.5dV
Service and Repair Manual
Number 022-780
Issue Number: 02
Page 13 of 32
10. Front Panel Replacement
43.
Remove the base plate as per 4, Instructions 1 to 5
NOTE:
General anti-static precautions
should be observed.
44.
Cut any cable ties securing the
looms together.
45.
Disconnect the Emergency Stop
loom (352-188) from the
Emergency Stop button on the front
panel to
E-STOP
(x8) on the VIC
board
46.
Disconnect the DV 20 way front
panel loom from the front panel.
47.
Remove the 6-off M3 x 16 pan head
screws (420-478).
48.
The front panel may now be
removed from the base casing.
Refitting of the Front Panel
is the reverse of the above procedure
Note:
When refitting the front panel, ensure that the self-adhesive EMC gasket
material is not disturbed. Replace damaged sections or the entire gasket if
necessary.
43
44
45