MultiTest
-dV
Operating Manual
Mecmesin
37
Limits
In a limit test, the crosshead cycles between the limit switches. To set your limit switches
loosen either screw, located on the rear of the test stand, and move the switch to the
required position.
A cycle starts when the crosshead is at the first limit switch and ends back at the same
limit switch. At the end of the test, the home button appears on the stand to enable the
crosshead to be moved back to the starting position.
Setting
Options
Cycle Count
0-9999
Up Speed
0-1200mm/min
Down Speed
0-1200mm/min
Start Direction
Choose whether the test direction is upward or downward.
Move to Start
Select if the test moves to the start position first.
Example
1.
Load your grips and sample,
2.
Set the upper limit switch at the desired upper displacement point,
3.
Set the lower limit switch at the desired lower displacement point,
4.
Set the desired start direction and whether to move to start,
5.
Configure the number of cycles,
6.
Execute the test. The stand moves between the two displacements for the configured
number of cycles and then stop.