Force Cycle
Up Speed:
100 mm/min
Down Speed:
20 mm/min
Start Direction:
Down
Test Sub-Type:
Force Cycle
Limit Force:
200 N
Return Force:
50 N
Cycle Count:
10
The crosshead moves down at 20 mm/min until the applied load is 200 N from tared
zero. Once the limit force is reached the crosshead moves upward at 100 mm/min until
a load of 50 N is reached, this cycle repeats 10 times at which point the test stops.
Break
Up Speed:
50 mm/min
Down Speed:
100 mm/min
Start Direction:
Up
Test Sub-Type:
Break
Break Threshold:
5 %
The crosshead moves up at 50 mm/min until a break condition is detected. The break
event must also occur above 5% of the connected AFG’s capacity.
Data Capture within VectorPro™
To use AFG Control/Force Control tests in conjunction with VectorPro first program
the test settings using the stand’s front panel and/or AFG and then create a VectorPro
test using the AFG/Force control operation, ensuring the speed and test direction match
your configuration.
Please note the speed and test direction use the settings configured within VectorPro,
while other test settings are loaded from the front panel. For more information please
refer to the VectorPro user manual, part no. 431-955.